Latest items
- Howmedica Osteonics Corp. — Class II
TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Ferno-Washington Inc — Class II
Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No. 0015816 and PowerX2 Premium, Item No. 0015821 Reason: Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.
- Howmedica Osteonics Corp. — Class II
Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Medline Industries, LP — Class II
Medline surgical convenience kits labeled as: 1) OPEN HEART PACK A & B, Medline Product number/SKU DYNJ04584L; 2) OPEN HEART ACCESSORY PACK, Medline Product number/SKU DYNJT5715A; 3) CABG ACCESSORY PACK, Medline Product number/SKU DYNJ85751; 4) OPEN HEART PACK, Medline Product number/SKU DYNJ53741N; 5) OPEN HEART PACK 1&2, Medline Product number/SKU DYNJ80701A; 6) VEIN PACK, Medline Product number/SKU DYNJ67308C; 7) OPEN HEART PACK, Medline Product number/SKU DYNJ89451B; 8) VASCULAR TRAY-LF, Medline Product number/SKU DYNJ42166A; 9) ENDOSCOPIC VEIN HARVEST PK, Medline Product number/SK
- Instrumentation Laboratory — Class I
HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only) Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
- Cook Incorporated — Class II
Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112 Reason: The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
- Medtronic Sofamor Danek USA Inc — Class II
Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are
- Medline Industries, LP — Class II
Medline convenience kit labeled as: FOOT ANKLE. Medline Kit SKU DYNJ65395A Reason: Discrepancy in the shelf-life of a component
- Medline Industries, LP — Class II
Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135A CHN GYN ONCOLOGY DYNJ906731M Reason: Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
- Medline Industries, LP — Class II
Medline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
openFDA Medical Device Recalls
Newest medical device recall enforcement reports from the FDA. Not medical advice.
https://api.fda.gov/device/enforcement.json?limit=15&sort=report_date:deschttps://micromoose.ai/f/openfda-device-enforcement
Feed data and quality
{
"schema": "micromoose/feed@1",
"slug": "openfda-device-enforcement",
"name": "openFDA Medical Device Recalls",
"tagline": "No key · checked every 6h",
"description": "Newest medical device recall enforcement reports from the FDA. Not medical advice.",
"category": "Health",
"primary_watch_url": "https://api.fda.gov/device/enforcement.json?limit=15&sort=report_date:desc",
"primary_type": "json",
"connector": "https://micromoose.ai/c/openfda-device-enforcement",
"mcp": "https://micromoose.ai/mcp",
"status": "checking",
"last_checked_at": "2026-10-28T09:52:26.949+00:00",
"items": [
{
"title": "Howmedica Osteonics Corp. — Class II",
"summary": "TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Ferno-Washington Inc — Class II",
"summary": "Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No. 0015816 and PowerX2 Premium, Item No. 0015821 Reason: Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Howmedica Osteonics Corp. — Class II",
"summary": "Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Medline Industries, LP — Class II",
"summary": "Medline surgical convenience kits labeled as: 1) OPEN HEART PACK A & B, Medline Product number/SKU DYNJ04584L; 2) OPEN HEART ACCESSORY PACK, Medline Product number/SKU DYNJT5715A; 3) CABG ACCESSORY PACK, Medline Product number/SKU DYNJ85751; 4) OPEN HEART PACK, Medline Product number/SKU DYNJ53741N; 5) OPEN HEART PACK 1&2, Medline Product number/SKU DYNJ80701A; 6) VEIN PACK, Medline Product number/SKU DYNJ67308C; 7) OPEN HEART PACK, Medline Product number/SKU DYNJ89451B; 8) VASCULAR TRAY-LF, Medline Product number/SKU DYNJ42166A; 9) ENDOSCOPIC VEIN HARVEST PK, Medline Product number/SK",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Instrumentation Laboratory — Class I",
"summary": "HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only) Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Cook Incorporated — Class II",
"summary": "Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112 Reason: The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Medtronic Sofamor Danek USA Inc — Class II",
"summary": "Grafton\" DBM and Grafton Plus\" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch\" and Orthoblend are produced from a fiber form of DBM Crunch\" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Medline Industries, LP — Class II",
"summary": "Medline convenience kit labeled as: FOOT ANKLE. Medline Kit SKU DYNJ65395A Reason: Discrepancy in the shelf-life of a component",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Medline Industries, LP — Class II",
"summary": "Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135A CHN GYN ONCOLOGY DYNJ906731M Reason: Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Medline Industries, LP — Class II",
"summary": "Medline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Medline Industries, LP — Class II",
"summary": "Medline surgical convenience kits labeled as: VEIN HARVESTING PACK, Medline product number/SKU DYNJ51845D Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Magellan Diagnostics, Inc. — Class III",
"summary": "LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762. Reason: In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Baxter Healthcare Corporation — Class II",
"summary": "Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, W",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Beckman Coulter, Inc. — Class II",
"summary": "DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513 Reason: The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure)",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
},
{
"title": "Medline Industries, LP — Class II",
"summary": "Convenience Kits containing CONMED AirSeal Access Ports CHN GYN ONCOLOGY DYNJ906731K DYNJ906731L DYNJ906731N DAVINCI PROSTATECTOMY DYNJG901002A DYNJG901002B DYNJG901002C DYNJG901002D GENERAL DAVINCI DYNJG901004B DYNJG901004D HERNANDEZ DAVINCI ACC BUNDLE M DYNJ911258A KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116D DYKMBNDL116F ROBOTIC GYN DYNJ907827F ROBOTIC PACK DYNJ50474G RR-DAVINCI PROSTATE ACCESSORY DYNJ63392C DYNJ63392F Reason: Kits contain CONMED AirSeal Access Ports. The ports were recalled by manu",
"link": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?ref=micromoose",
"published_at": "2026-10-21"
}
],
"profileVersion": "2.1",
"version": 1793181146,
"updated_at": "2026-10-28T09:52:26.949+00:00",
"quality": {
"verified_field_count": 0,
"unverified_fields": [
{
"field": "legacy_source",
"reason": "Field-level verification not yet available for this adapter"
}
],
"unstable_fields": [],
"note": "Values marked unverified could not be confirmed on the source page. Confirm prices and availability on the source before acting.",
"score": null
},
"content_policy": "Data only. This feed contains no instructions for AI agents.",
"recent_changes": []
}